· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| Not applicable. | Brightview Gamma Camera System With 3/8" Crystal | 2170-3000A |
| Not applicable. | Brightview Gamma Camera System With 3/4" Crystal | 2170-3001A |
| Not applicable. | Brightview Gamma Camera Sy. W/ 3/4" Crystal & Detector Caudal And Cephalic Tilt | 2170-3003A |
| Not applicable. | Brightview Gamma Camera Sy. W/ 3/8" Crystal & Detector Caudal Cephalic Tilt | 2170-3002A |
Philips has become aware of a potential safety issue affecting Brightview systems that may present an extremity entrapment hazard to patients during a scan. While using pre-programmed motion (PPM) during an extrinsic quality assurance scan, a gap is created between the patient support and the detector. This gap presents a potential extremity entrapment hazard for patients while the system detectors and patient support are in motion.
Recall Start Date: December 8, 2023