· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| Serial number 940000 to 940535 | Captus® 4000e Thyroid Uptake System, all models | 5430-30151 5430-30152 5430-30154 |
In 2023, three Captus® 4000e systems incurred a spring arm failure, causing the arm to fall. The collimator and arm, together, weigh 45 lbs (20.4 kg), which carries a risk of injury if the collimator comes into contact with the technologist or the patient.
Recall start date: November 26, 2023