Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Azurion 7 M20
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.Azurion 7 M20722 224

Issue

Philips has become aware of two potential safety issues with the Azurion 7 M20 Flexarm System ("flexarm") with a Maquet Magnus Table with software version 05.02.12 ("table"). Due to communication issues between the flexarm and the table, motorized movements of the table and/or the flexarm may be lost. There can be a connection issue (software issue with table to be corrected in field action FCO72200557). There is also a date and time issue where when the flexarm system is (re)started, the flexarm and table may not connect due to a software issue. This occurs when the date/time format includes more than three (3) leading zeros (to be corrected in field action FCO72200528).

Recall start date: December 13, 2023