Health Products

Azurion 7 M20

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Azurion 7 M20
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.Azurion 7 M20722 224

Issue

Philips has become aware of a potential safety issue with the Azurion 7 M20 Flexarm System ("flexarm") with a Maquet Magnus Table with software versions 05.02.12 and 05.02.15 ("table"). Due to a communication issue between the flexarm and the table, motorized movements of the table and/or the flexarm may be lost. A representative of Getinge (manufacturer of the Maquet Magnus Table) will arrange a site visit to return the software of your table to version 05.02.12. After this change, your system will be subject to the medical device recall letter (HS23039R/2023-IGT-BST-020).

Recall start date: December 13, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime