· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| Not applicable. | Azurion 7 M12 | 722 223 722 078 |
| Not applicable. | Azurion 5 M12 | 722 227 |
| Not applicable. | Azurion 5 M20 | 722228 |
| Not applicable. | Azurion 7 B12 | 722 067 |
| Not applicable. | Azurion 7 M20 | 722 224 722 079 |
| Not applicable. | Azurion 7 B20 | 722 226 722 068 |
Philips has become aware of a potential safety issue with the Philips Azurion Systems R1.0, R1.1, R1.2, R2.0, R2.1, and R2.2, where the system may exhibit a loss of imaging functionality and data. Philips has identified a potential issue which can cause the system to continuously restart (restart loop). T his is likely to occur when the patient database gets too big over time (>500 studies). If this issue occurs, a potential for data loss may be expected.
Recall start date: December 15, 2023