Health Products

Azurion 7 System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Azurion 7 System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Not applicable.Azurion 7 M12722 223
722 078
Not applicable.Azurion 5 M12722 227
Not applicable.Azurion 5 M20722228
Not applicable.Azurion 7 B12722 067
Not applicable.Azurion 7 M20722 224
722 079
Not applicable.Azurion 7 B20722 226
722 068

Issue

Philips has become aware of a potential safety issue with the Philips Azurion Systems R1.0, R1.1, R1.2, R2.0, R2.1, and R2.2, where the system may exhibit a loss of imaging functionality and data. Philips has identified a potential issue which can cause the system to continuously restart (restart loop). T his is likely to occur when the patient database gets too big over time (>500 studies). If this issue occurs, a potential for data loss may be expected.

Recall start date: December 15, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime