Health ProductsType I recall

Philips Achieva, Intera and Ingenia Systems

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Philips Achieva, Intera and Ingenia Systems
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Model or catalog number/Lot or serial numberAffected products
3.0TNot applicable.Ingenia 3.0T
INTERA 1.5TNot applicable.Intera 1.5T MRI System-Main Unit
ACHIEVA 1.5TNot applicable.Achieva 1.5T MRI System-Main Unit
ACHIEVA XRNot applicable.Achieva XR System-Main Unit
ACHIEVA 3.0TNot applicable.Achieva 3.0T MRI System-Main Unit

Issue

The quadrature body coil (QBC) seal adhesive may fail creating sharp edges that may come in contact with patients. The QBC seal may become loose as the patient table travels in a horizontal motion in and out of the system bore. The QBC seal (figure 1) is a rubber seal that is glued between the cone cover and QBC cover and functions to prevent sharp edges of the QBC cover from contacting patients during an examination.

Recall start date: January 9, 2024