Medfusion® Syringe Infusion Pumps

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Medfusion® Syringe Infusion Pumps
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Contact the manufacturer.Medfusion® Syringe Infusion Pumps3500
3500-415
3500-500
Contact the manufacturer.Medfusion® Model 4000 Syringe Infusion Pump4000-0105-51
4000-0106-01
4000-0101-51

Issue

Smiths Medical identified that certain barrel clamp guides may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded. The affected barrel clamp guide part (mechanism) number is G6000716 and the associated affected lot numbers are P0407365, P0486670 and P0561740. The listed Medfusion® syringe infusion pumps, contain the barrel clamp guide mechanism.

Recall start date: December 28, 2023