· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| D2D0262RY D2E1492RY D2E1210RY D2E1375RY D1K1916FRY D2J0691RY D1M0153RY D2E1491RY D2E2348RY | Sofsilk Nonabsorbable Braided Silk Surgical Suture | SS675 S246 S1750K S318 S605 S608 S1172 S1789K S243 |
| D2C3250FRY | Ti-Cron Reverse Cutting 3/8 Circle | 8886307051 |
| D1M2840RY D2C0774RY | Ti-Cron Taper Cardiopoint, 300 Series Steel 1/2 Circle | 8886308731 8886318541 |
Specific lots of the Sofsilk™ Braided Silk sutures and Ti-Cron™ Coated Braided Polyester Sutures were sterilized with gamma doses that exceeded the range approved or were exposed to more than the approved number of Ethylene Oxide (EO) sterilization cycles. The affected lots were distributed between May 2022 and November 2023. The extra gamma levels or additional EO cycle have the potential to decrease the strength of these sutures over time.
Recall Start Date: January 25, 2024