Health Products

Alissa Interpret

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Alissa Interpret
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
v5.3.7
v5.4.3
Alissa InterpretK5849AA

Issue

1. The ACMG (American College of Medical Genetics and Genomics) classification scheme implemented in Alissa Interpret does not result in variant of uncertain significance (VOUS) classification when benign and pathogenic variant are contradictory. 2. The ACMG classification scheme does not result in the correct classification when the criteria strength is updated.

Recall Start Date: December 21, 2023