Health Products

Servo-N

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Servo-N
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
v. 4.4 or 4.5 with HFO-optionServo-N6688600

Issue

A limitation of the intended performance of Servo-N during high frequency oscillatory ventilation (HFOV) when using the Fisher & Paykel Neonatal Circuit (950N81*) and the humidifier FP950. The use of a relatively small endotracheal (ET) tube in relation to body weight and in combination with the Fisher & Paykel Neonatal Circuit* that is used with the FP950 humidifier limits the delivered Pampl and therefore reduces the delivered VThf.

Recall Start Date: January 22, 2024