· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| 23BBM531 | Composite Neck Dissection SHSC-LF | DYNJ46776D |
| 23BBL719 23BBE584 23CBR156 | Parotid | DYNJ905111A |
| 23FBH342 | Simple Neck Dissection | DYNJ47481F |
| 23BBJ268 23BBK356 23ABC131 22LBR066 23CBS913 | Pacemaker PK SHSC-LF | DYNJ46785C |
Specific lots of the Sofsilk™ Braided Silk Sutures and Ti-Cron™ Coated Braided Polyester Sutures were sterilized with gamma doses that exceeded the range approved or were exposed to more than the approved number of ethylene oxide (EO) sterilization cycles. The affected devices were distributed as part of Medline packs between Sep 2022 and January 2024. The extra gamma levels or additional ethylene oxide (EO) cycle have the potential to decrease the strength of these sutures over time.
Recall start date: January 19, 2024