· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots. | Airflow™ Small Adult Resuscitator (AF5000 Series) | AF5140MBS AF5100MB |
| All lots. | Horizon™ Adult Resuscitator | HZI9111 |
| All lots. | Airflow™ Pediatric Resuscitator (AF2000 Series) | AF2140MB-PW AF2100MB |
| All lots. | Airflow™ Infant Resuscitator (AF3000 Series) | AF3100MB AF3100MBP AF3100MB-M1 |
| All lots. | Airflow™ Small Child Resuscitator (AF4000 Series) | AF4100MBP |
| All lots. | SafeSpot™ Infant Resuscitator (SS3200 Series) | SS3200MB SS3200MB-2 |
Sunmed Holdings LLC (also trading as Ventlab LLC) is updating its IFU for the manual resuscitators to include a warning statement. A safety bulletin was issued relative to the use of accessories (e.g., peep valve) on the exhalation port of the manual resuscitator which may increase the backward leak due to increased expiratory resistance and that monitoring of end tidal carbon dioxide is recommended in critically ill patients.
Recall start date: October 11, 2023