Health Products

Atlan® A350

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Atlan® A350
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Model or catalog number/Lot or serial numberAffected products
8621600ASRJ-0212
ASNN-0052
Atlan® A350 XL
8621500More than 10 numbers, contact manufacturer.Atlan® A350

Issue

The recall is being initiated as the market surveillance activities revealed a few cases in which the internal backup battery failed spontaneously while the Atlan® is being operated without mains supply. This resulted in an unexpected shutdown of the device while it was running on batteries. The affected Atlans started with a battery status of 100%, but quickly shut down and did not trigger the alarm for low battery. A secondary acoustic alarm signal which is independent from mains and battery power was generated as specified. No patient consequences have been reported to Draeger so far.

Recall start date: November 21, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime