Care Plus Incubator

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Obstetrics and gynaecology
Product: Care Plus Incubator
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
More than 10 numbers, contact manufacturer.3000
4000
1000
Care Plus Incubator

Issue

GE Healthcare has become aware that the bedside panel of the incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.

Recall start date: February 5, 2024