Health Products

SOLTIVE™ Premium SuperPulsed TF Laser System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: SOLTIVE™ Premium SuperPulsed TF Laser System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.SOLTIVE™ Premium SuperPulsed TF Laser SystemTFL-PLS

Issue

As part of Olympus' post market surveillance of SOLTIVE™ laser use, including a review of customer complaints and adverse events, we have become aware that some clinicians are defaulting to the preset laser values for all clinical procedures when that is not consistent with the instructions for use. Olympus is reminding users that preset treatment parameters may not be appropriate for all patients and are guidelines only. Olympus is sending this reminder after reviewing twenty-eight (28) complaints reported between 2021 to 2023 related to ureteral stone procedures, that described serious injuries.

Recall start date: February 5, 2024