· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| More than 10 numbers, contact manufacturer. | 106524 106524US | Heartmate 3 LVAS Implant Kit |
| More than 10 numbers, contact manufacturer. | 103695-CAN 106015 | Heartmate II LVAS - Implant Kit |
| More than 10 numbers, contact manufacturer. | 104911 | Heartmate II LVAS - Implant Kit With Sealed Grafts |
Abbott is notifying customers of a planned update to instructions for use associated with observed outflow graft deformation known as "extrinsic outflow graft obstruction" (EOGO) associated with the Heartmate 3™ left ventricular assist system (LVAS) and Heartmate II™ LVAS. Significant EOGO will manifest clinically as as a persistent low flow alarm under certain circumstances in some patients, and in such cases, may impair the ability of the Heartmate LVAS to provide adequate hemodynamic support.
Recall start date: February 20, 2024