· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| 230922-018009 | 84055060D | Origio Sperm Wash |
During testing it was found that one of the batches, had an insufficient concentration of sodium pyruvate - only 0.002 mg/ml compared to a specified concentration of 0.046-0.137 mg/ml (1.5-4.3% of the specified amount). Part of this formulation batch had also been used to create three lots of commercially released sperm wash medium.
Recall start date: January 22, 2024