| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| More than 10 numbers, contact manufacturer. | Auto Suture Blunt Tip Trocar | OMS-T10BTS OMS-T10BTNL OMS-T10BTSNL OMS-T12BT OMS-T10BT OMS-T12BTNL |
| More than 10 numbers, contact manufacturer. | Auto Suture Structural Balloon Trocar | OMS-T10SB |
Medtronic received reports of events related to Covidien Auto Suture™ Structural Balloon Trocar and Auto Suture™ Blunt Tip Trocar reporting trocar seal disengagement when using the trocar with certain mesh products. This issue can occur when the trocar is used to deploy the mesh not in accordance with the mesh instructions for use (IFU). Not following the IFU for the mesh including use of incompatible trocar size, excess force, and not hydrating the mesh and/or folding of the mesh could result in a seal disengaging into the cavity.
Recall Start Date: March 4, 2024