Health Products

Perseus A500

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Perseus A500
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.Perseus A500MK06000

Issue

The recall is being initiated as the market surveillance activities revealed a few cases in which the internal backup battery failed spontaneously while the perseus device being operated without mains supply. This resulted in an unexpected shutdown of the device while it was running on batteries. The affected devices started with a battery status of 100%, but quickly shut down and did not trigger the alarm for low battery. A secondary acoustic alarm signal which is independent from mains and battery power was generated as specified. No patient consequences have been reported to Draeger so far.

Recall Start Date: March 6, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime