Health Products

Medfusion Syringe Infusion Pumps

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Medfusion Syringe Infusion Pumps
Issue: Medical devices - Performance
What to do: Contact JENI PANDYA at 514‐905‐2012 (JENI.PANDYA@ICUMED.COM) and SHUI YEE KUAN at 514‐905‐2645
Audience: Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Not applicable.Medfusion Syringe Infusion Pumps3500G
3500SD-500
3500VX-500
3500-402
3500E
3500VX-415
3500
3500-415
3500VX-414
3500-306
3500-414
3500-500
3500SD
3500VX
3500ZE
3500BC
3500VX-306
Not applicable.Medfusion Model 4000 Syringe Infusion Pump4000-0101-249
4000-0101-50
4000-0105-51
4000-0105-249
4000-0105-78
4000-0106-00
4000-0106-01
4000-0101-78
4000-0100-50
4000-0106-231
4000-0101-51
4000-0105-50

Issue

Smiths Medical is issuing this letter to notify customers not to use Cardinal Health branded Monoject syringes with Medfusion 3500 and 4000 Syringe Pumps. On September 20, 2023, Cardinal Health issued a medical device product correction for the Cardinal Health branded Monoject Luer-Lock Tip Syringes due to demonstrated recognition and compatibility issues with syringe infusion pumps due to changes made to the dimensions of the syringes. As a result, Cardinal Health recommends that these syringes not be used on syringe infusion pumps.

Recall Start Date: February 29, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime