· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| More than 10 numbers, contact manufacturer. | LISS/Coombs | 004014 004017 |
| More than 10 numbers, contact manufacturer. | Coombs Anti-IgG | 004024 004026 004027 |
| More than 10 numbers, contact manufacturer. | ID-DiaCell I-II-III | 004310 |
An increased level of complaints and adverse events related to weak non-specific reactions when using ID-system screening reagents in Indirect Antiglobulin Testing (IAT) for qc, donor, and patient samples. Customers using both, the manual and instrument methods reported weak positive reaction results (identified as "?", "wr", "+/-", and sometimes "+" on instruments) instead of an expected clearly negative "-" reaction in IAT. These doubtful reactions or in some cases weak positive reactions may occur randomly among batches. The investigations so far point out that a possible enhancer of the phenomenon is the combination of certain lots of reagents red blood cells with certain lots of associated ID-cards.
Recall start date: March 11, 2024