· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| More than 10 numbers, contact manufacturer. | 2843241 2843181 2843191 | Catheter Extension |
The manufacturer determined through testing that non-dioxin-like polychlorinated biphenyl acids (PCBAs) could leach from the stay•safe catheter extension, which are manufactured using peroxide cross-linked silicone tubing. Please note that PCBAs are chemically different from polychlorinated biphenyls (PCBs), which were not identified. To date, no complaints have been received related to PCBAs. PCBA levels decrease with shorter stay•safe catheter extension and decrease over the treatment time.
Recall start date: February 20, 2024