Health Products

Hugo Ras Tower, 120v

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: Hugo Ras Tower, 120v
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
C22AKC0070MRASC0003Hugo Ras Tower, 120v

Issue

There have been 22 complaints of reported communication errors with the Hugo™ Ras System. Medtronic determined that the errors were caused by specific ports on a network switch located on the back of the Hugo™ tower. The communication errors can present in a variety of ways and may be intermittent, but the system always alerts the user of an error. In some cases, the communication errors may require a system reboot.

Recall start date: March 18, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime