Health Products

Cardiohelp-I

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Gastroenterology and urology
Product: Cardiohelp-I
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
90410131
90410130
Cardiohelp-I70104.8012

Issue

During service measurements on the Cardiohelp-I devices in the field, it was noticed that the measured patient leakage current exceeded the limit values of 50 μa specified in standard EN 60601-1:2006 + a1:2013. Patient leakage currents of up to 1337 μa were measured. Investigations have shown that only Cardiohelp-I devices with installed sensor panels (701051308) with an old sensor bridge (701049193) hardware version are affected. The new hardware version is not affected.

Recall Start Date: March 18, 2024