· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| Not applicable. | 866113 M3150 M3150B M3140 866112 M3145 866094 M3155 866091 866093 | Intellivue -Information Centre |
The ups devices are backup batteries utilized by PIC IX central monitoring stations and their associated hardware components. These devices are intended to provide power to the PIC IX system or components (monitors, networking devices, servers, etc.) in the event of power outages ensuring that the PIC IX system remains functional during power fluctuations. It has been discovered that an internal component of the ups devices produces less output energy than specified, resulting in the ups' inability to power up as required, which affects the power supply to PIC IX system. These failures can occur during a power outage or when the ups is only acting as a conduit for the sites main line power to the PIC IX system.
Recall start date: March 28, 2024