· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| version 8.1.0.47 version 12.0.0.932 version 10.0.1.52 version 7.0.0 version 12.0.4.12 version 10.1.0.613 version 12.1.0.1221 version 10.1.1.54 version 8.1.2.5 | 11B SP3 10A SP1 10B SP1 11B SP1 | RayStation |
An issue was found with the calculation of the reported SSD (i.e., the SSD displayed and exported) in RayStation RayPlan 7-2024a including some service packs. The reported SSD may in very rare cases be too high. The dose calculation is not affected by this issue.
Recall start date: February 23, 2024