Health Products

iMRI 3T S

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: iMRI 3T S
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
10093-01iMRI 3T S116118-000

Issue

The recall is being initiated to address a notification received by Deerfield Imaging, Inc. from Siemens regarding potential dot- or line-shaped artifacts caused by the presence of lubricating grease within the headband of the over-ear headphones when used during head examinations with the Magnetom MRI systems. Siemens became aware of this issue during post market surveillance. Siemens has identified that the root cause of the issue was a manufacturing processing error.

Recall Start date: April 12, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime