Health Products

Patient Information Center iX

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Patient Information Center iX
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Not applicable.Patient Information Center iXVERSION 4.2
VERSION 4.0
VERSION 4.1

Issue

It has been discovered that when the user copies the customized event categories from one unit to another, the event catalog displays the customized event categories, but the changes are not actually applied. If a user reviews the events in the event monitor the user will see in the event details that the event category is not updated to reflect the customization. This issue is the result of cached changes to the event catalog not propagating in the database.

Recall Start Date: April 17, 2024