Health Products

Vercise Genus R16 and R32 IPG Kits

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Neurology
Product: Vercise Genus R16 and R32 IPG Kits
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
100209 - 753347Vercise Genus R16 IPG KitDB-1216
100104 - 753200Vercise Genus R32 IPG KitDB-1232

Issue

Boston Scientific has received reports of undesired sensation and/or transient worsening of movement disorder symptoms secondary to a transient loss of stimulation during IPG charging for patients with rechargeable DBS IPGs. investigation of these reports determined that the device may reset (by design) due to detection of noise or interference during IPG charging.

Recall Start Date: April 18, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime