· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots. | BiPAP A40 Pro, CA W/ Heated Tube Humidifier, CA | CAX3100T12 |
| All lots. | BiPAP A40 Pro, Ca | CAX3100S12 |
| All lots. | BiPAP A30 System | 1111147 |
| All lots. | BiPAP A30 System - Ventilator & System One A-Series Heated Humidifier | 1111148 |
| All lots. | BiPAP A40, Canada | 1109596 1111173 1111174 |
The affected products feature a ventilator inoperative alarm, which occurs when the ventilator detects an internal error or a condition that may affect therapy. this may manifest in the following ways:
The device may reboot intermittently for 5-10 seconds (stops providing therapy, screen goes blank during the reboot, and there is a single audible alert), restarting therapy, and returning to delivering therapy with same patient settings.
or
The device may reboot intermittently for 5-10 seconds (stops providing therapy, screen goes blank during the reboot, and there is a single audible alert), restarting therapy, and returning to delivering therapy but with factory default settings.
or
When there are three (3) reboots within a 24-hour period, the device will enter a ventilator inoperative state (therapy stopped, audible and visual alarms present).
or
The device may enter a ventilator inoperative state without a reboot preceding this condition.
Any of the above scenarios could result in interruption and/or loss of therapy which may lead to hypoventilation, mild to severe hypoxemia, hypercarbia, respiratory failure/insufficiency, or potentially death in the most vulnerable patients.
Recall start date: March 19, 2024