Health ProductsType I recall

EsoFLIP Dilation Catheter 30mm

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: EsoFLIP Dilation Catheter 30mm
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected products/Model or catalog number
23D0777JZ
22K0746JZ
EsoFLIP Dilation Catheter 30mmES-330

Issue

Medtronic has received complaints of ES-330 catheters measuring diameters that are significantly (more than 50%) lower than actual. The failure mode may be reported as "failed precheck", "incorrect distensibility reading", "no distensibility value displayed", "adverse event without identified device or use problem", or "diameter incorrect." Medtronic Lafayette returned product analysis lab testing of unused, returned ES-330 catheters showed saline conductivity that is approximately 10 times lower than specification. The ES-330 saline conductivity specification is 2.25 mS/cm. The saline conductivity of multiple unused, returned ES-330 catheters had saline conductivity at or near 0.25 mS/cm.

Recall start date: April 25, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime