Alaris™ Pump Infusion Set Low Sorbing Tubing (PE Lined) Back Check Valve 2 SmartSite™ Y-site(s)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Alaris™ Pump Infusion Set Low Sorbing Tubing (PE Lined) Back Check Valve 2 SmartSite™ Y-site(s)
Issue: Medical devices - Performance
What to do: Contact Becton Dickinson Canada Inc.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
22115368Alaris™ Pump Infusion Set Low Sorbing Tubing (Pe Lined) Back Check Valve 2 SmartSite™ Y-site(s)10013072

Issue

BD has received customer reports of the Alaris™ Pump Infusion Set Low Sorbing Tubing (PE Lined) drip chamber becoming detached from the tubing. BD has confirmed that this issue is isolated to catalog number 10013072 only and a single batch (no. 22115368) where mis-assembly of the drip chamber to the tubing occurred. All other active catalog numbers containing a drip chamber are not affected.

Recall start date: April 18, 2024