· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| Serial number 860791 | Mi1250 SYNCHRONY 2 Flex (26) | 39575 |
| Serial number 863442 Serial number 859126 | Mi1250 SYNCHRONY 2 Flex (28) | 39573 |
MED-EL has recently identified a small number of devices, which may have a higher probability of a crack within the internal electronics substrate which could result in a non-functional device. MED-EL wants to avoid the implantation of any defective device and voluntarily request their immediate return from clinics and distributors.
Recall start date: April 19, 2024