· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| 19029603AA0124001 19029603AA0224008 19029603AA0124002 | IOMAX Cortical Module | 190296-200 |
An assembly error was discovered on the cortical module (PN 190296-200). A cortical module insulator that is used to provide an electric barrier between circuit boards for TCS-9, ICS, and other internal parts was found to be installed incorrectly. The insulator provides clearance between safety critical components and incorrect installation does not meet the minimum creepage/air gap required to meet IEC 60601-1.
Recall start date: April 23, 2024