· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots. | Luer Slip Disposable Syringe Without Needle | SYR101020 |
| All lots. | Syringe, Luer Lock | SYR101010 |
It has been identified that testing documentation does not support that the syringes can successfully deliver accurate volumes across the full range of their claimed graduated capacity and may pose a risk to patient health. Syringes not being able to deliver the intended dose may cause inaccurate dosing leading to the need for additional treatment or other medication.
Recall start date: April 29, 2024