Philips Hemodynamic Application

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Philips Hemodynamic Application
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Not applicable.Philips Hemodynamic Application002010

Issue

Philips has become aware of a potential issue with Philips Interventional Hemodynamic Application R1.2.X, R1.3.0 and R1.3.1, where synchronization of pressure wave data received from two devices, the Intellivue X3 and Functional Measurement Patient Interface Module (FM-PIM) may not be correct. There might be a delay of the FM-PIM data with respect to the Intellivue X3 data which is not constant in time. When this occurs, the calculation of measurements will be affected, specifically the IFR (Instantaneous Wave-Free Ratio) and FFR (Fractional Flow Reserve) values.

Recall Start Date: May 3, 2024