· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots. | InstaClear Sheath | LCS4K00UNOL LCS1830BTST LCS1500UNOL LCS4K45TPOL LCS4K70BTOL LCS1800UNST LCS4K70TPOL LCS4K30TPOL LCS1570TPOL LCS4K30BTOL LCS1530TPOL LCS1530BTOL LCS1830TPST |
Olympus (Gyrus Acmi, Inc.) is initiating a medical device recall action after investigating a complaint for an InstaClear Sheath (model # LCS4K30BTOL) in which the plate on the distal end of the sheath broke off during a procedure. Olympus has received sixty-seven (67) complaints across all sheath models related to breakage or damage to the pin or plate at the distal end of the InstaClear Sheath since the product was launched in 2016. Four (4) complaints were reported as serious injuries.
Recall Start Date: May 3, 2024