· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| Serial number 23023421 Serial number 23063829 Serial number 22113113 Serial number 23063830 Serial number 22123270 Serial number 23023429 | NovaGuide™ 2 Intelligent Ultrasound System | NSC-NVGSYS2-CA |
| Serial number 22074378 Serial number 23033578 Serial number 22103022 Serial number 23063856 Serial number 21123067 Serial number 22064204 Serial number 22074379 Serial number 22064205 Serial number 21123068 Serial number 23104174 | NovaGuide™ 2 Traditional Ultrasound System | NSC-TCDNG2 |
A potential for an unintentional disconnection of the internal chasis ground cable was identified. This was initially addressed on 02/16/24, with a final design on 04/29/24. The cable is not accessible during normal operation and the inadvertant disconnection is a remote possibility.
Recall start date: May 6, 2024