Fabian™ HFO Ventilator

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Fabian™ HFO Ventilator
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.Fabian™ HFO Ventilator112001
113001

Issue

Software V5.2.2 changed the criteria for detecting a patient circuit disconnect during HFO ventilation with the intent to reduce the potential for patient safety-related risks.

While ventilating in HFO mode with pressure amplitude (Pamp) > 1.5x mean pressure (Pmean) and with the breathing circuit properly connected, if the breath-by-breath leak monitor exceeds 40%, the ventilator triggers the disconnect alarm. the displayed volume and leak monitors reset to zero and the ventilator reduces the HFOV amplitude as a protection mechanism to prevent lung injury upon reconnection following a true disconnection. If the patient is not disconnected, this results in a small increase in Pmean, and potentially large reduction in amplitude until the alarm is cleared.

The potential health risk for a false disconnect alarm occurring that cannot be resolved is inadequate ventilation due to the oscillation being out of range, resulting in hypoxia, hypercapnia, hypoventilation and/or respiratory arrest.

Recall start date: April 17, 2024