Health Products

BVI Safety Knife

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Ophthalmology
Product: BVI Safety Knife
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
6058854378227BVI Safety Knife

Issue

Manufacturer received feedback from a singular customer that the wrong configuration of the blade was in the package. The bevel was on the opposite side, instead of being in front (bevel up) it was in the back (bevel down). There was no injuries or illnesses reported in connection to the associated failure mode.

Inadvertently, using a bevel down blade instead of a bevel-up blade causes the surgeon to make an unintended incision architecture. A careful, observant surgeon is likely to notice the incorrect blade orientation under the microscope prior to use. However, when unnoticed, the unintended incision architecture potentially leads to an incision leakage. When incision leakage goes unnoticed , a watertight wound closure may not be achieved and thereby increasing the risk for anterior chamber fluid egress or ocular surface fluid ingress. Health effects include infection, iris prolapse, and intraocular pressure (IOP) fluctuations.

Recall start date: May 10, 2024

Get told about the next health product recall.