Health Products

MobileDiagnost wDR

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: MobileDiagnost wDR
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productModel or catalogue number
More than 10 numbers, contact manufacturer.MobileDiagnost wDR9890 010 8952X

Issue

Philips has been made aware by supplier Sedecal of a potential problem related to interference between the steel cable and metal plate in the carriage of MobileDiagnost wDR systems' telescopic arm. The potential interference is caused by the shape of the cable's crimped terminals. This issue may result in failure of the anti-fall system which would prevent the telescopic arm from falling if the steel cable breaks. If the steel cable breaks and the anti-fall system does not activate, the telescopic arm may fall on operators, patients, or service personnel. Philips has not received any reports of this issue or harm as of April 2024.

Recall start date: May 17, 2024