· Published by Health Canada · Official notice
| Lot or serial number | Affected product | Model or catalogue number |
|---|---|---|
| More than 10 numbers, contact manufacturer. | Coons Dilator-Polyethylene | JCD12.0-38-20-COONS JCD14.0-38-20-COONS JCD10.0-38-20-COONS |
| More than 10 numbers, contact manufacturer. | Vessel Dilator-Polyethylene | JCD 8.0-38-20 JCD 8.0-38-15 JCD10.0-38-20 JCD 8.0-38-20-HC JCD20.0-38-20 JCD22.0-38-20 JCD16.0-38-20-HC JCD14.0-38-20-HC |
| More than 10 numbers, contact manufacturer. | Disposable Chib Biopsy Needles | DCHN-18-15.0 DCHN-22-10.0 DCHN-22-20.0 DCHN-18-10.0 DCHN-22-10.0-U DCHN-22-15.0 DCHN-22-15.0-U |
| More than 10 numbers, contact manufacturer. | Peel-Away Introducer Sets And Trays | PLVW-XXX |
Cook Medical identified that devices from the affected device lots may have packaging seals that do not meet peel strength specifications. This creates a potential of a sterile barrier failure that could result in compromised device sterility.
Recall start date: March 1, 2024