· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| AY19G008474 | Z Vent | Z VENT |
| AY20A009538 AY20A009669 AY16J001624 AY20A009546 | Uni-Vent 731 EMV+ | 731 EMV+ |
| VCCC100515 AY17F002444 VCDC120895 AY17C002076 VCDC120882 | Eagle II Portable Ventilator | 731 EAGLE II |
Manufacturer is voluntarily recalling a limited number of operator's guides and quick reference guides related to Zoll ventilators with MRI compatibility. Through an internal review of product manuals, it was identified that documented patient safety ventilator MRI information was inadvertently omitted from the manual. The impacted manual only states to place the ventilator behind the 2000 gauss line. The updated language has been changed to: You must place the ventilator behind the 130 gauss field line (approximately 2 meters to the bore opening of a 3T MRI magnet). This update to the instruction ensures safe distance from the MRI to ensure proper function of the ventilator and reduce the risk of delay in ventilator therapy.
Recall Start Date: May 24, 2024