Health Products

NGENUITY 3D Visualization System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: NGENUITY 3D Visualization System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Software v1.5NGENUITY 3D Visualization System8065000399

Issue

Alcon has recently become aware of an anomaly within the NGENUITY™ surgical image guidance functionality available in software version 1.5 (NGENUITY™ ID 8065000399). The anomaly can lead to inaccurate rotational positioning of the image guidance overlay. This anomaly only affects the surgical image guidance functionality within software v1.5. To mitigate the potential for this anomaly, users will be advised to re-perform registration prior to ending the case or if any axis discrepancy is noted.

Recall Start Date: May 23, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime