Health Products

DELFIA / AutoDELFIA PIGF 1-2-3

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: DELFIA / AutoDELFIA PIGF 1-2-3
Issue: Medical devices - Performance
What to do: Contact the importer REVVITY HEALTH SCIENCES CANADA, INC. at TURKUQMRESPONSE@REVVITY.COM.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.DELFIA / AutoDELFIA PIGF 1-2-3B055-301

Issue

We have become aware through customer complaints that the measured PlGF analyte concentrations in part of the patient samples (approximately 15%) may be decreased ≥ 50%. The decreased PlGF concentration results may cause raise in the amount of high risk/false positive pre-eclampsia and Down's syndrome screening results.

Recall Start Date: May 29, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime