· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| Not applicable. | Achieva 1.5T MRI System - Sense Torso Coil | 453567141882 453567141883 4535 675 02881 |
| Not applicable. | Achieva XR System - Sense Coil | 453567141882 453567141883 453567394942 |
| Not applicable. | Achieva 3.0T MRI System - Sense Torso Coil | 4535 673 94942 4535-673-9494X |
| Not applicable. | Intera 1.5T MRI System - Sense XL Torso Coil | 453567141882 453567141883 |
Philips has been made aware through customer complaints of a potential safety issue with specific SENSE XL Torso (1.5T and 3.0T) coils that could pose a risk for patients. Philips has identified an issue in the SENSE XL Torso (1.5T and 3.0T) coils which may result in localized heating during the scan, resulting in potential harm to patients. If the patient is exposed to localized heating there is a potential for a heating sensation and/or 1st, 2nd, or 3rd degree burns around the area covered by the coil.
Recall start date: June 6, 2024