· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| Lot and Batch # 024.0000786207 Lot and Batch # 024.0000791526 Lot and Batch # 024.0000813195 Lot and Batch # 024.0000806730 Lot and Batch # 024.0000798137 | FL12-60 | Flowpra Class I And II Screening Tests |
A suspected false negative for HLA class I with FL1-30 lot 022. Complaint investigation testing was performed using 4 samples with low levels of A2 antibody specificities and 1 sample with low levels of A31 antibody specificities. Investigation confirmed a lower percentage of positive reactions for samples tested with FL1-30 lot 022. In addition, investigation noted that lot specific user document, TDX-OLI-DMR-LOT-8536 revision 01, has incorrect reference values for the distribution of certain HLA class I antigen specificities.
Recall start date: May 29, 2024