Health Products

Flowpra Class I And II Screening Tests

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Haematology
Product: Flowpra Class I And II Screening Tests
Issue: Medical devices - Performance
What to do: Instructions provided on response.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
Lot and Batch # 024.0000786207
Lot and Batch # 024.0000791526
Lot and Batch # 024.0000813195
Lot and Batch # 024.0000806730
Lot and Batch # 024.0000798137
FL12-60Flowpra Class I And II Screening Tests

Issue

A suspected false negative for HLA class I with FL1-30 lot 022. Complaint investigation testing was performed using 4 samples with low levels of A2 antibody specificities and 1 sample with low levels of A31 antibody specificities. Investigation confirmed a lower percentage of positive reactions for samples tested with FL1-30 lot 022. In addition, investigation noted that lot specific user document, TDX-OLI-DMR-LOT-8536 revision 01, has incorrect reference values for the distribution of certain HLA class I antigen specificities.

Recall start date: May 29, 2024