| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| G71467 G77408 G73624 G82557 G79899 G82499 | SM-40HF-B-D-VIR | Easy Moving Plus B Mobile X-Ray Unit With Digital Detector |
Notification to inform customers regarding potential safety issue related to the steel cable installed inside the mobile column which supports the weight of telescopic arm with the tube head. The potential risk consists of an interference which could cause that the safety system which blocks the arm of the cable is cut off does not work properly. The interference occurs between the cable and a metal plate of the arm's carriage due to shape of the cable's crimped terminals. Scope of the safety recall is to review the installed base according to the manufacturer's instructions and install the kit immediately if the cable could interfere with the parachute system.
Recall start date: June 5, 2024