· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| More than 10 numbers, contact manufacturer. | HORIZON C HORIZON A HORIZON W | Horizon X-Ray Bone Densitometer (Dxa) |
During standard safety design verification tests, Hologic has identified a non-conformance in Horizon DXA devices. The non-conformance pertains to electromagnetic compatibility requirements according to the international technical standard IEC 60601 - 1 - 2 for the safety and essential performance of medical electrical equipment, where the result from the Horizon DXA system exceeded the limit for acceptability.
Recall start date: June 14, 2024