Health Products

Infinix and Alphenix Systen C-Arm Suspension and Table

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Infinix and Alphenix Systen C-Arm Suspension and Table
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
Not applicable.INFX-8000C/PF
INFX-8000C/QU
Alphenix Infx-8000c-Main Unit
Not applicable.INFX-8000V_/M4
INFX-8000V/K2
INFX-8000V_/M5
Infinix Cf-I/Sp-System
Not applicable.INFX-8000H/KE
INFX-8000H/K5
INFX-8000C_/MT
INFX-8000C_/ME
INFX-8000C/KF
INFX-8000C/K6
INFX-8000C_/MF
Infinix Vc-I-System
Not applicable.CAS-830AC-Arm Support
Not applicable.INFX-8000V_/MDInfinix Cf-1/Bp-System
Not applicable.INFX-8000V/K3
INFX-8000V_/M7
INFX-8000V_/ML
Infinix Vf-I/Sp-System
Not applicable.INFX-8000V/PU
INFX-8000V/PV
INFX-8000V/P5
INFX-8000V/PW
Alphenix Infx-8000v-Main Unit
Not applicable.INFX-8000V_/MFInfinix Vf-I/Bp-System
Not applicable.INFX-8000C_/M5
INFX-8000H/K5
Infinix Cc-I-System

Issue

It has been discovered that the affixing screws of the C-arm motor drive and table, that are part of the x-ray interventional system, may become loose. As a result, abnormal noises may be generated from the C-arm suspension and/or the table. Also, sensor errors may be displayed. However, please note that the affixing parts will not fall nor fail.

Recall start date: June 7, 2024