· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| Not applicable. | INFX-8000C/PF INFX-8000C/QU | Alphenix Infx-8000c-Main Unit |
| Not applicable. | INFX-8000V_/M4 INFX-8000V/K2 INFX-8000V_/M5 | Infinix Cf-I/Sp-System |
| Not applicable. | INFX-8000H/KE INFX-8000H/K5 INFX-8000C_/MT INFX-8000C_/ME INFX-8000C/KF INFX-8000C/K6 INFX-8000C_/MF | Infinix Vc-I-System |
| Not applicable. | CAS-830A | C-Arm Support |
| Not applicable. | INFX-8000V_/MD | Infinix Cf-1/Bp-System |
| Not applicable. | INFX-8000V/K3 INFX-8000V_/M7 INFX-8000V_/ML | Infinix Vf-I/Sp-System |
| Not applicable. | INFX-8000V/PU INFX-8000V/PV INFX-8000V/P5 INFX-8000V/PW | Alphenix Infx-8000v-Main Unit |
| Not applicable. | INFX-8000V_/MF | Infinix Vf-I/Bp-System |
| Not applicable. | INFX-8000C_/M5 INFX-8000H/K5 | Infinix Cc-I-System |
It has been discovered that the affixing screws of the C-arm motor drive and table, that are part of the x-ray interventional system, may become loose. As a result, abnormal noises may be generated from the C-arm suspension and/or the table. Also, sensor errors may be displayed. However, please note that the affixing parts will not fall nor fail.
Recall start date: June 7, 2024